Table 2. Comparison of sensor offsets before and after temperature calibration on August 13, 2024

Calibration Time O1x (nT) O1y (nT) O1z (nT) O2x (nT) O2y (nT) O2z (nT) O3x (nT) O3y (nT) O3z (nT)
Calibrated (CAL) 00:00:00 71.3 –19.3 6.0 –57.8 –27.4 114.4 –0.4 3.3 101.1
08:00:00 71.5 –19.2 6.1 –57.9 –25.4 121.5 –0.9 4.1 101.6
16:00:00 72.3 –19.2 5.2 –57.4 –25.4 120.2 –0.6 3.8 100.1
Temperature-calibrated (TCAL) 00:00:00 69.5 –19.4 6.0 –58.9 –25.3 115.6 –0.4 3.3 101.2
08:00:00 69.4 –19.4 6.1 –59.1 –22.8 122.8 –1.2 4.0 101.8
16:00:00 70.1 –19.4 5.3 –58.7 –22.7 121.5 –1.0 3.7 100.3